Antegrade Femoral Lengthening with a Magnetic Intramedullary Nail

A magnetic telescopic nail allows gradual femoral distraction after osteotomy using an external controller. This guide covers antegrade insertion for femoral shortening. The proximal entry point and the osteotomy level are separate decisions. The entry point follows the nail design and proximal femoral anatomy. Osteotomy level depends on the deformity, canal geometry, and the implant support available at the end of lengthening. [1,2]

Assoc. Prof. Mehmet Selcuk SAYGILI · Pendik Bolge Hospital, Istanbul
Last updated: Nov 16, 2026

Indications and Approach Selection

Consider an antegrade magnetic nail when the femoral component of a limb-length discrepancy warrants surgical correction and the canal can accommodate the implant. Assess hip and knee motion, skeletal maturity, bone quality, prior infection or surgery, and the patient’s ability to follow the distraction, rehabilitation, and weight-bearing plan. Check that the proposed implant is locally authorized for the intended use. [1,3]

Antegrade entry is an option for isolated femoral shortening. If a major distal femoral deformity also requires correction, assess whether retrograde insertion better matches the planned osteotomy and fixation. Proximal deformity or retained hardware may require a different implant or reconstruction strategy. [1,2]

Preoperative Planning

  • Measure length and alignment. Use calibrated full-length standing radiographs when feasible. Determine the femoral contribution to the discrepancy, assess the mechanical axis, and review coronal and sagittal femoral geometry. Document rotation clinically and obtain a torsional study when it would change the correction plan. [1,2]

  • Template the construct at the start and end of lengthening. Check nail diameter, length, and available stroke. Confirm that the larger nail segment will still adequately engage the distal fragment at the planned final length and that both interlocking zones have sufficient bone. [1]

  • Match the entry to the nail. A proximally bent trochanteric-entry nail and a straight piriformis-entry nail have different trajectories. Confirm a canal-centered path in both planes. In skeletally immature patients, assess the physes and model-specific labeling. Avoid piriformis entry in the immature proximal femur because of the risk to femoral head perfusion. [1,3,4]

Surgical Technique

  1. Position and image. Use a radiolucent setup that allows orthogonal views of the hip, planned osteotomy, and distal femur. Mark rotational references before completing the osteotomy. [1]

  2. Prepare the entry and canal. Confirm the starting point and guidewire trajectory in both planes. Prepare the canal for the selected implant according to its instructions. Reassess if the guidewire is eccentric or the nail meets unexpected resistance. [1,4]

  3. Complete the planned osteotomy. Use a low-energy technique with limited periosteal disruption. Advance the nail across the osteotomy under imaging while maintaining control of the fragments. Check for fracture propagation and confirm alignment and rotation before locking. [1]

  4. Lock and verify. Use proximal and distal interlocking according to the implant design. If the canal is wide at the osteotomy and the nail does not adequately control alignment, consider blocking screws based on the anticipated direction of drift. Verify alignment, hardware position, and distraction function according to the device instructions. Mark the magnet location for controller placement when the system requires it. [1,4]

Distraction and Follow-up

Set the latency, rate, and rhythm at discharge. Review serial radiographs to compare actual distraction with the prescribed amount and assess regenerate formation and alignment. Adjust the prescription under the treating team's supervision if regenerate formation is poor, consolidation is premature, or pain and loss of motion limit rehabilitation. [1]

Start supervised rehabilitation early, with attention to hip extension and knee flexion. Use the selected device's weight-bearing limits and radiographic evidence of consolidation to guide progression. At follow-up, document limb length, coronal and sagittal alignment, hip and knee motion, and any new neurologic symptoms. Discuss removal timing at the outset and follow the current instructions for the implanted model. [1,3]

Pitfalls

Pitfall

What to check

Varus or procurvatum

Verify entry and proximal fragment alignment in both planes. Reassess stability if the canal is wider than the nail at the osteotomy. [1]

Unintended rotation

Compare the pre-osteotomy rotational references before final locking unless a correction is planned. [1]

Inadequate support at final length

Template segment engagement at the intended endpoint before choosing nail length and osteotomy level. [1,4]

Loss of joint motion or delayed regenerate

Track motion and regenerate throughout distraction and adjust the plan promptly when either deteriorates. [1,5]

Implant overload or distraction failure

Confirm the model-specific loading limits. If radiographic distraction differs from the prescription, check controller placement, implant function, and the osteotomy before continuing. [1,3,5]

Evidence and Counseling

A systematic review found complications in 332 of 983 lengthened segments (34%) across studies of PRECICE and FITBONE. The estimate combines devices and anatomical segments and reflects heterogeneous reporting. Discuss device failure, delayed or premature consolidation, contracture, and the possible need for additional surgery. [5]

OrthoRico Take-home Message: Match the entry point to the nail, template the construct at the final length, and check alignment and joint motion throughout distraction.

References

  1. Rozbruch SR, Fragomen AT. Lengthening of the femur with a remote-controlled magnetic intramedullary nail: antegrade technique. JBJS Essent Surg Tech. 2016;6(1):e2.

  2. Fragomen AT, Rozbruch SR. Lengthening of the femur with a remote-controlled magnetic intramedullary nail: retrograde technique. JBJS Essent Surg Tech. 2016;6(2):e20.

  3. US Food and Drug Administration. Update: NuVasive Specialized Orthopedics’ Precice Devices. 2023.

  4. NuVasive Specialized Orthopedics. PRECICE Antegrade and Retrograde Femur Surgical Technique Guide. Archived technical reference.

  5. Frost MW, Rahbek O, Traerup J, Ceccotti AA, Kold S. Systematic review of complications with externally controlled motorized intramedullary bone lengthening nails (FITBONE and PRECICE) in 983 segments. Acta Orthop. 2021;92(1):120–127.